kevin_nust 发表于 2008-7-23 09:28

请教信号采集卡的问题

学生在调bently的ADRE408的时候,通过信号发生器产生了一个峰峰值为4,正电压偏置为10V的正弦信号来模拟加速度信号,结果信号采集卡上的指示灯显示:“Channel input is over or under range”。请问各位老师问题出在什么地方?
望能得到各位老师的指点,谢谢!
信号采集卡的参数:
Channels
8
A/D Converters
24 bit, S/N ratio of 111 db typical
Input Signal Coupling
AC or DC
Static Variables (each channel)
Direct Amp, Bandpass Amp, 1X and 2X Amp and Phase, Average and Instantaneous Gap, Multiple Speed Values and Date/Time Stamp
Waveform Variables (each channel)
4 user-defined Synchronous or Asynchronous waveforms plus a raw waveform
Connectors
SMA SubMiniature version A. A small threaded coax cable connector. - Sub Miniature A
Analog Front End Conditioning

Transducer Input
positive or negative bias
Input Impedance
Single Ended and Differential: 742 kOhms
4-20 mA: 511 Ohms
Upper and Lower Input Voltage
User-configurable
Full Scale Range
User-configurable
Transducer Scale Factor
User-configurable

w89986581 发表于 2008-7-23 09:34

可能是你设置的10V的直流偏置导致信号超出采集卡的量程了。

Galaxy163 发表于 2008-7-23 09:52

为什么要设置10V的正电压偏置?

kevin_nust 发表于 2008-7-23 11:16

回复 3楼 的帖子

师兄说要设置一个直流偏置的,我也不知道哎,郁闷的哈

kevin_nust 发表于 2008-7-23 11:17

回复 2楼 的帖子

我去掉,试试看:lol :lol

alien016 发表于 2008-7-23 14:59

回复 楼主 的帖子

是直流偏置电压设置过大,超出了采集卡的测量范围了!通常采集范围为-10V~+10V,在你设置偏置以后,你信号发生器的输出电压为8V~12V,所以会出现信号灯警告!
输出的信号的最高电压值不能超过采集卡的极限范围!:@)

kevin_nust 发表于 2008-7-25 22:21

回复 6楼 的帖子

正如您所说的,问题已经解决了,谢谢大家:lol

xautmcf 发表于 2008-7-26 09:42

回复!!

可能超出测量范围了!!建议扩大量程再试一ixia
页: [1]
查看完整版本: 请教信号采集卡的问题